Detail of Recall

Type of Product:Device
Medsafe Reference:21164
Brand Name:LIFEPAK 20 and LIFEPAK 15 Manual Defibrillator
Model:
Affected:LIFEPAK 15 Monitor/Defibrillator with End-Tidal CO2 option.
Software version:
Recalling Organisation:Medxus Ltd
Unit 4
100 Fitgerald Avenue
Christchurch 8011
Contact Information:03 365 7403
Manufacturer:Physio-Control Corporation
Issue:During the investigation the manufacturer of the EtCO2 module found that compressed air may have contaminated filters within in the EtCO2 module. As a result of the contamination of the filter, the EtCO2 display on the LIFEPAK 15 will provide a reading of XXX. There may be solid or dashed lines present on the device display in the place of an EtCO2 waveform and the SERVICE LED will be illuminated. In this situation the EtCO2 monitoring functionality will no longer be available.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/3/2017