Detail of Recall

Type of Product:Device
Medsafe Reference:21161
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model:
Affected:Artis Q systems with A100G generators
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue: In Artis systems with A100G generators, a component can be affected by aging, resulting in the failure of a module in the high-voltage generator. When the failure occurs, the system usually responds by providing emergency fluoroscopy. In the worst case, a spontaneous failure of the radiation generation function is possible.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/2/2017