Detail of Recall

Type of Product:Device
Medsafe Reference:21124
Brand Name:Brightview XCT
Model:
Affected:Software Issues:2.5.1, 2.5.2, 2.5.3, 2.5.4 and 2.5.5
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer has become aware of a potential problem with BrightView XCT, with software Issues in 2.5.5 and earlier.
Issue 1 - Motion Controller emit the warning message
Occasionally, the system may stop and you may receive a "XCT scan did not complete successfully".
Issue 2 - Door interlock switch with X Ray perform
System was unable to perform an X-Ray scan, after the system displays a "XCT Door Interlock Error.
Issue 3 - Detector contacts head holder when performing Patient Unload
When performing a PreProgrammed Motion (PPM) to unload the patient, the detector will contact the head holder under the following circumstances:
- Head holder installed
- Rel-180 detector orientation
- Gantry Rotate angle between 45° and 55°
Issue 4 - JETStream Freezes during Gated Planar scan
When acquiring a Gated Planar scan (MUGA - Multi Gated Acquisition) on a patient with an irregular heartbeat, you may encounter a situation where the acquisition station locks at 100% complete
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/2/2017