Detail of Recall

Type of Product:Device
Medsafe Reference:21117
Brand Name:ID-DiaCell l-ll-lll ID-DiaPanel
Model:
Affected:All lots currently in date
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Brand Name:ID-DiaCell ABO
Model:
Affected:ID-DiaCell IP-IIP-IIIP, ID-DiaScreen VP-VIP, ID-DiaPanel-P, All lots currently in date
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:The manufacturer has become aware of unexpected reactions primarily on eluates and QC samples but also with some patient samples when using the ID-DIACell l-ll-lll and ID-DiaPanel products. In case of doubtful reactions, please re-test with a new kit. If the results remain in doubt, the manufacturer advises to send the sample to a reference laboratory and if an urgent transfusion is required, perform a crossmatch.
Follow up 24 Feb 2017: There has been further development (follow up) from our Manufacturer additional 3 other devices that are also affected.
The manufacturer received complaints regarding indirect antiglobulin test (IAT), especially when eluates and QC samples were tested, less frequently when patient samples were tested.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/2/2017