Detail of Recall

Type of Product:Device
Medsafe Reference:21111
Brand Name:Abbott i-STAT1 Analyzer Software
Model:
Affected:JAMS142E
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Point of Care Canada Ltd
Brand Name:Abbott Celite ACT Cartridges
Model:
Affected:03P86-25
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Point of Care Canada Ltd
Brand Name:Abbott Kaolin ACT Cartridges
Model:
Affected:03P87-25
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Point of Care Canada Ltd
Issue:The manufacturer has determined that Celite ACT and Kaolin ACT cartridges may produce falsely low results when very high clotting time samples are tested. Patient samples with a measured clot time of >853 seconds for Celite and >1134 seconds for Kaolin may display results of approximately 100 to 200 seconds while using JAMS142E. The typical therapeutic range of the i-STAT ACT assay is 200 to 600 seconds. Results obtained within this range are unaffected.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/2/2017