Detail of Recall

Type of Product:Device
Medsafe Reference:21092
Brand Name:Agfa Computed Radiography NX Workstation
Model:
Affected:Software 3.0.8950
Software version:
Recalling Organisation:Cass Medical Limited
803 Paparata Road
Bombay
SOUTH AUCKLAND
Contact Information:0800 66 77 80
Manufacturer:Agfa-Gevaert NV
Issue:Due to a memory issue a warning message about unexpected error and possible incorrect data integrity is shown to the user. A reboot of the workstation computer after having received the warning message prevents the issue from occurring. In one instance the warning instructions were not followed and resulted in the image of one patient being archived in another patient's record.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/2/2017