Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21083 |
| Brand Name: | Angiojet ULTRA CONSOLE and JetStream CONSOLE |
| Model: | |
| Affected: | Console foot |
| Software version: | |
| Recalling Organisation: | Boston Scientific NZ Ltd 3 Diamond Street Eden Terrace Auckland 1021 |
| Contact Information: | 0800 742 678 |
| Manufacturer: | Boston Scientific Corporation |
| Issue: | The manufacturer has received field reports regarding the AngioJet Ultra 5000A Console foot switch sticking during use resulting in an undesired continued operation of the device. This situation can occur as a result of the wear and tear of the foot switch, which is normally captured through routine servicing. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/2/2017 |