Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21075 |
| Brand Name: | Beckman Coulter iChemVELOCITY Automated Urine Chemistry Systems |
| Model: | |
| Affected: | 800-3530, 800-7167, 800-7719, 800-7720, 800-3564, 800-7100, 800-7103, 800-7106, 800-7162, 800-7163, 800-7166, 800-3061 |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Iris International |
| Issue: | The manufacturer has determined that sample probe misalignment or bending on the iChemVELOCITY may lead to a remote possibility of false negative results due to inadequate strip dosing. The investigation was initiated based on customer complaints for control failures. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/2/2017 |