Detail of Recall

Type of Product:Device
Medsafe Reference:21072
Brand Name:Roche Cobas 4800
Model:
Affected:cobas HBV Test for use on the cobas 4800 system (06979564190)
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:In extremely rare circumstances, the algorithm for the cobas® HIV-1 and HBV Tests for use on the cobas® 4800 system may inappropriately call an extraordinarily high viral load sample as negative (i.e., false negative result). This situation occurs when the positive PCR growth curve for the target has a Ct value below the Ct minimum for the test, resulting in the kinetic algorithm calling the curve negative instead of invalid.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/2/2017