Detail of Recall

Type of Product:Device
Medsafe Reference:21056
Brand Name:Radiometer ABL800 series blood gas analysers
Model:
Affected:All connected to HIS/LIS or middleware Data Management system
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:On the ABL800 series analyzers the sample type cord blood has been changed into the two sample types cord blood arterial and cord blood venous. This change took place with software version 3.1 MR5. An operator or clinician who is aware that the sample types that can be selected on ABL800 series analyzer are cord blood arterial and cord blood venous may notice that a less specific sample type cord blood appears in the HIS/LIS or middleware system.
Followup April 2017: The final solution to this issue is an upgrade of the ABL800 software to software version 6.17 or higher. This new version ensures that cord blood arterial and cord blood venous are transmitted correctly to a HIS/LIS or a middleware system.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/4/2017