Detail of Recall

Type of Product:Device
Medsafe Reference:21046
Brand Name:RaySearch RayStation Treatment Planning System
Model:4.5, 4.7, 5, 6 and 2
Affected:
Software version:
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:Investigation showed that in a certain, non-standard, remoting installation, accelerated computation on GPU may be wrong for a very large computation problem. This can affect deformable registration and dose computation.
The error affectes optimization and there is a risk that the error drives the optimizer into creating a plan that apears acceptable although there is unacceptable dose in some region.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/1/2017