Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21046 |
| Brand Name: | RaySearch RayStation Treatment Planning System |
| Model: | 4.5, 4.7, 5, 6 and 2 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | Investigation showed that in a certain, non-standard, remoting installation, accelerated computation on GPU may be wrong for a very large computation problem. This can affect deformable registration and dose computation. The error affectes optimization and there is a risk that the error drives the optimizer into creating a plan that apears acceptable although there is unacceptable dose in some region. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/1/2017 |