Detail of Recall

Type of Product:Device
Medsafe Reference:21041
Brand Name:Siemens Sysmex CS-2000i, CS-2100i, CS-2500 and CS-5100 Systems
Model:
Affected: 10488064, 10488062, 11239235 and 10709128 respectively.
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed that under very rare circumstances a false negative result may be reportable. The issue may only occur under the following conditions:
1. The initial result is falsely flagged with an antigen excess error that triggers an automatic reanalysis in a 1/19 dilution.
2. The raw signal that is generated from the reanalysis is below the Limit of Blank (LoB) of the method.
Follow-up communication
The manufacturer has issued a new protocol disc version with new and enhanced assay applications for Sysmex CS-2000i/CS-2100i Systems. This also includes a new protocol disc version with new and enhanced assay applications for Sysmex CS-2500 Systems and on Sysmex CS-5100 systems.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory