Detail of Recall

Type of Product:Device
Medsafe Reference:21035
Brand Name:Symmetry Surgical - Sterilization Container
Model:
Affected:Ultra System Sterilization Container - C13136
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Symmetry Surgical Inc
Issue:The manufacturer has become aware that the device may not maintain its contents in a completely sterile condition as intended post-sterilisation. Subsequent investigation and testing have disclosed an observation that a gap in the device's base trim needs to be inspected to verify that sufficient sealant exists to ensure that the container seal is maintained.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/1/2017