Detail of Recall

Type of Product:Device
Medsafe Reference:21025
Brand Name:Siemens Calibrator A
Model:
Affected: Catalogue number 10285903
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, the manufacturer confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/1/2017