Detail of Recall

Type of Product:Device
Medsafe Reference:21018
Brand Name:Philips PIIC iX Central System
Model:
Affected:Cisco 3850 layer 3 switch
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:When the InfiniBand cable is used with the Cisco 3850 switch, the interface is set to err-disable when the cable is inserted. The port will then not work properly which is detected during Test and Inspection at the time of installation.
If the issue occurs after installation, for example if the cable is disconnected/reconnected or if the switch reboots, a loss of traffic will occur. As a result, your monitoring solution may experience a loss of centralized monitoring.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/1/2017