Detail of Recall

Type of Product:Device
Medsafe Reference:21016
Brand Name:Siemens Mammomat Fusion/Inspiration
Model:
Affected:10140000 with software version VB30 and with a Biopsy option
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:The manufacturer has become aware of potential errors that may occur while performing a biopsy. The system may freeze and the biopsy functionality may become not available in the following situations:
a) Drop of the compression force, e.g. this may occur due to a patient sudden movement during the active targeting phase.
b) Or use of the inversion tool while the scout image is active during the targeting phase
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/1/2017