| Type of Product: | Device |
| Medsafe Reference: | 21005 |
| Brand Name: | B. Braun Electrosurgical system generator, general-purpose |
| Model: | |
| Affected: | GN200 |
| Software version: | |
| Recalling Organisation: | B Braun New Zealand Ltd HSBC Tower 188 Quay Street Parnell AUCKLAND 1010 |
| Contact Information: | 021 784 724 |
| Manufacturer: | Aesculap AG |
| Issue: | In exceptional cases the devices might have a functional disorder, which will be detected by the device itself and signaled to the user unmistakably by way of an error message (ERROR F008, Code 26) and an acoustic alarm. The generator will be set into a safe mode in which no power will then be released. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/1/2017 |