Detail of Recall

Type of Product:Device
Medsafe Reference:20998
Brand Name:Hsiner_Manual Resuscitator (Disposable)
Model:
Affected:Product Code: HS-60122 and HS-60123. Batch/Lot/Serial numbers: 20160926019, 20160926020, 20161108023 and 20161108066
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Hsiner Co Ltd
Issue:The manufacturer has identified the possibility of a functional failure on the Pop-Off Valve assembled on the disposable manual resuscitator. The valve may have a dimension issue, resulting in the valve being unable to fully close off when the patient needs higher pressure (40cm H2O and above).
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/1/2017