Type of Product:Device
Medsafe Reference:20997
Brand Name:Monaco
Model:
Affected:Running Monaco: V 5.10.00 (including 5.10.01, 5.10.02), or V 5.11.00 (including 5.11.01), or V 5.20.00, and Using Elekta Motorized Wedges
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:The manufacturer has indentified that calculated dose from wedged beams can be assigned to the wrong beam. This can occur for a plan with at least two beams using the Elekta motorized wedge. The user must change the beam index numbers belonging to the two wedged beams, such that the index numbers have been exchanged for these beams, then edit the wedge angle of one of the beams.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/1/2017