Detail of Recall

Type of Product:Device
Medsafe Reference:20996
Brand Name:Siemens ADVIA Centaur Systems, Dimension Vista Systems, IMMULITE Systems
Model:
Affected:ADVIA Centaur Progesterone1 - All Lots, Dimension Vista LOCI Progesterone2 - All Lots, IMMULITE/IMMULITE 1000 Progesterone - All Lots, IMMULITE 2000 Progesterone3 - All Lots
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed that the presence of DHEA-S (a metabolite of DHEA, a steroid hormone that may be used as part of in vitro fertilization (IVF) protocols to improve ovarian response and IVF treatment outcomes) causes falsely elevated progesterone results
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/1/2017