Type of Product:Device
Medsafe Reference:20995
Brand Name:Ortho Clinical Diagnostics Vitros Chemistry Product Na Slides
Model:
Affected:Generation (GEN) 1 through 6, 30 through 40, 44 through 49
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics
c/o Quigg Partners
Level 7, 36 Brandon Street
Wellington 6011
Contact Information:0800 402 899
Manufacturer:Ortho-Clinical Diagnostics Inc
Issue:The manufacturer has identified the potential for biased results to be generated over the 10 day on-analyzer limit when using VITROS Na+ Slide cartridges warmed between 1 1/2 to 8 hours then placed on the analyzer. The manufacturer has determined that cartridges require a minimum of 8 hours of warm up at room temperature to help ensure the slides have acceptable performance throughout the 10 day on-analyzer storage limit.
Follow up and resolution November 2017
The manufacturer is introducing VITROS Electrolyte Reference Fluid (ERF) 300 and VITROS Electrolyte Reference Fluid (ERF) 800 with a revised formulation that:
1. Restores the Na+ cartridge warm-up protocol to 90 minutes when taken from the refrigerator (120 minutes from the freezer) as defined in the VITROS Na+ Slides Instructions for Use.
2. Reformulated ERF helps to reduce sources of variability (e.g., calibration variability) observed when using VITROS Na+ Slides.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/11/2017