Detail of Recall

Type of Product:Device
Medsafe Reference:20994
Brand Name:Light handle system
Model:
Affected:NS-3600-B LITE GLOVE 1000/CASE, 3611 FLEXBL LITE GLOVE 1EA/PKG, 3613 LITE GLV-FLEXIBLE 3EA/PKG, 3612 LITE GLV-FLEXIBLE 2EA/PKG
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Covidien LLC
Issue:Customers have reported that, on rare occasion, the Devon Light Glove may split upon application to the Devon Light Handle Adapter. The splits resulted from difficult application of the Light Glove to the Handle Adapter. More recently, clinicians have reported finding splits in the Light Glove following surgery completion.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/1/2017