Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20994 |
| Brand Name: | Light handle system |
| Model: | |
| Affected: | NS-3600-B LITE GLOVE 1000/CASE, 3611 FLEXBL LITE GLOVE 1EA/PKG, 3613 LITE GLV-FLEXIBLE 3EA/PKG, 3612 LITE GLV-FLEXIBLE 2EA/PKG |
| Software version: | |
| Recalling Organisation: | Covidien NZ Ltd Level 3, Building 5, Central Park Corporate Centre 666 Great South Road Penrose AUCKLAND 1051 |
| Contact Information: | 0508 489 264 |
| Manufacturer: | Covidien LLC |
| Issue: | Customers have reported that, on rare occasion, the Devon Light Glove may split upon application to the Devon Light Handle Adapter. The splits resulted from difficult application of the Light Glove to the Handle Adapter. More recently, clinicians have reported finding splits in the Light Glove following surgery completion. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/1/2017 |