Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20993 |
| Brand Name: | Plum 360 Infusion pump |
| Model: | |
| Affected: | Software Version 15.02 |
| Software version: | |
| Recalling Organisation: | Hospira NZ Ltd 58 Richard Pearse Drive Airport Oaks Mangere Auckland 2022 |
| Contact Information: | 04 384 7463 |
| Manufacturer: | Hospira Inc |
| Issue: | The manufacturer identified two software defects with the Plum 360 Infusion Pump, Software Version 15.02 resulting in the following issues: 1. The pump software incorrectly generates a Depleted Battery alarm instead of a Replace Battery alarm when the battery has a near end of life charge capacity 2. When AC power is lost and quickly resumed, the pump software incorrectly classifies the event as an E323 alarm condition. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/1/2017 |