Detail of Recall

Type of Product:Device
Medsafe Reference:20993
Brand Name:Plum 360 Infusion pump
Model:
Affected: Software Version 15.02
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Issue:The manufacturer identified two software defects with the Plum 360 Infusion Pump, Software Version 15.02 resulting in the following issues:
1. The pump software incorrectly generates a Depleted Battery alarm instead of a Replace Battery alarm when the battery has a near end of life charge capacity
2. When AC power is lost and quickly resumed, the pump software incorrectly classifies the event as an E323 alarm condition.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/1/2017