Detail of Recall

Type of Product:Device
Medsafe Reference:20984
Brand Name:Roche Diagnostics Cobas c501/502 XN-9000
Model:
Affected:XN-9000 complete with XN-10/20 on Version 20 Build 8
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer has been notified of a software bug identified on the XN-9000 line products in which patient information was replaced by that of another patient resulting in incorrect patient demographics being assigned to the patient sample and the results reported.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/12/2016