Detail of Recall

Type of Product:Device
Medsafe Reference:20973
Brand Name:TCVS Heart lung machine
Model:
Affected:Base for Terumo Advanced Perfusion System 1
Software version:
Recalling Organisation:Terumo Australia Pty Ltd (NZ)
c/- Healthcare Logistics
56 Carrington Road
Mount Albert
Auckland 1025
Contact Information:0800 66 77 57
Manufacturer:Terumo Cardiovascular Systems Corp
Issue:The manufacturer has received reports of instances in which the 6 Roller Pump for the Terumo Advanced Perfusion System 1 displayed a pump jam message and stopped while in use. Investigation concluded that, while there is no specific evidence that the 6 roller pumps were operating outside manufacturing specifications:
1. Pump jam is more likely to occur when using the 6 roller pump for low-RPM cardioplegia delivery than when using the 4 roller pump for dual-sized tube set cardioplegia delivery.
2. The occlusion settings suggested in the Operator's Manual for some dual-sized tubing sets may increase the likelihood of pump jam on the 6 roller pump.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/12/2016