Detail of Recall

Type of Product:Device
Medsafe Reference:20972
Brand Name:Siemens Dimension Vista V-LYTE Intergrated Multisensor
Model:
Affected:SMN 10484429
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed a low bias for Dimension Vista V-LYTE Integrated Multisensor urine potassium (K) when compared to the Dimension QuikLYTE Integrated Multisensor, which is the predicate device used in the method comparison section of the V-LYTE Instructions for Use (IFU). Our testing showed a bias of -4.9% at a urine potassium concentration of 39.6 mmol/L, -19.6% at a urine potassium concentration of 32.8 mmol/L, and an average bias of -10.9% over the urine potassium concentration range of 7.5 to 133.7 mmol/L. This bias affects patient results, QC and could result in failures in accuracy-based proficiency testing program.
The manufacturer is issuing this follow-up notification with enhanced instructions for updating and verifying the urine potassium method comparison coefficients.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/2/2017