Detail of Recall

Type of Product:Device
Medsafe Reference:20970
Brand Name:UniCel DxH 800 Coulter Cellular Analysis System and UniCel DxH 600 Coulter Cellular Analysis System
Model:
Affected:All Software Versions
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood samples. This is due to limitations in the available technology as well as sample limitations.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/12/2016