Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20968 |
| Brand Name: | Elekta Monaco |
| Model: | |
| Affected: | V 5.00.00 (including 5.00.01, 5.00.02, 5.00.03, 5.00.04), or V 5.10.00 (including 5.10.01, 5.10.02, 5.10.04), or . V 5.11.00 (including 5.11.01), or V 5.20.00, and Using the Collapsed Cone Algorithm, and . Using Varian Enhanced Dynamic Wedges (EDW) or Si |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | EDW or VW wedge files can be loaded into incorrect energies using Treatment Unit (TU) Storing. No warning or consistency checking is done for EDW or VW files when adding these to the database. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/12/2016 |