Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20965
Brand Name:Synthes_Fixation system implantation kit, internal [44759]
Model:
Affected:RIA Tube Assembly, for RIA Drive Shaft minimum length 520 mm, for No. 314.743, sterile
Software version:
Recalling Organisation:Johnson & Johnson (NZ) Limited
13a Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Synthes GmbH
Issue:The expiration date on the label for the product is incorrect. Existing testing supports an expiration date of 2 years from manufacturing. The affected products on the market were labelled with an expiration date of 10 years. There is biocompatibility test data to support an expiration date 2 years only; biocompatibility testing was not completed for 10 years
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/12/2016