Detail of Recall

Type of Product:Device
Medsafe Reference:20952
Brand Name:Siemens Artis Q, Q.zen, zeego, zee systems
Model:
Affected: In Artis systems with A100Plus or A100G generators
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:This corrective action addresses 2 possible, mutually independent causes of a system defect. In Artis systems with A100Plus or A100G generators, an attempt to resume operation following detection of a fault (such as a short circuit in the X-ray tube) can result in the failure of a module in the high-voltage generator.
For biplane systems delivered with software version VD11B since April of this year, software problems in conjunction with the graphics card may in rare cases result in the loss of image display in the examination room
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/12/2016