Detail of Recall

Type of Product:Device
Medsafe Reference:20941
Brand Name:Alaris Infusion Pumps
Model:
Affected:Models 8000 and 8015
Software version:
Recalling Organisation:CareFusion New Zealand 313 Limited
Unit 4 / 110 Mays Road
Onehunga
AUCKLAND 1061
Contact Information:0508 422 734
Manufacturer:CareFusion 303 Inc
Issue:The manufacturer has received reports of the Alaris System PC units operating on battery power where the low battery alarm and / or the very low battery alarm are not being triggered before the battery is discharged and all infusion channels are stopped. This issue does not occur if the device is plugged into an AC outlet.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/12/2016