Detail of Recall

Type of Product:Device
Medsafe Reference:20912
Brand Name:Abbott ARCHITECT Active-B12 Reagent kits
Model:
Affected:Lot number - 10533UP00, 10534UP00
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Diagnostics Division
Issue:Manufacturer has identified a lot specific stability issue with both reagent lots 10533UP00 and 10534UP00. As the reagent ages, Relative Light Unit (RLU) values for the calibration curve will decrease causing patient results calculated from this depressed curve to over-recover, potentially resulting in falsely elevated patient results. Controls may not reflect reagent instability as recalibration may result in control values within acceptable ranges.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/12/2016