Detail of Recall

Type of Product:Device
Medsafe Reference:20911
Brand Name:Getinge Group Heater Cooler Unit
Model:
Affected:Part Numbers: HU35
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Maquet Cardio Pulmonary GmbH
Issue:Investigations have confirmed that the current published disinfection procedure of the HU 35 system's water paths (in the Instructions for Use) does not adequately meet contemporary hygienic requirements and effectiveness against atypical mycobacteria.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/12/2016