Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20908 |
| Brand Name: | Philips Ingenuity CT |
| Model: | |
| Affected: | All systems supported by UPS devices. |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | Without preventive maintenance, and/or periodic battery replacement, UPS devices may fail, resulting in battery acid leakage, overheating, and/or the emission of fumes. All lead/acid battery-based UPS devices are susceptible to this issue. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/12/2016 |