Detail of Recall

Type of Product:Device
Medsafe Reference:20908
Brand Name:Philips Ingenuity CT
Model:
Affected:All systems supported by UPS devices.
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Without preventive maintenance, and/or periodic battery replacement, UPS devices may fail, resulting in battery acid leakage, overheating, and/or the emission of fumes. All lead/acid battery-based UPS devices are susceptible to this issue.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/12/2016