Type of Product:Device
Medsafe Reference:20892
Brand Name:Blood Donor Set/Quad/46347
Model:
Affected:filter type LCRD2
Software version:
Recalling Organisation:MacoPharma Australia Pty Ltd
BDO Auckland
L8
120 Albert St
Auckland 1010
Contact Information:
Manufacturer:Maco Pharma SA
Issue:The manufacturer was informed of a number of fast filtrations on the reference NPT6280LA. This has caused an absence of effective filtration through the media of the LCRD2 filter. This defect is linked to a partial absence of the weld in the media caused by poor positioning (fold or gap) of the last polyester layer in the mattress.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/12/2016