Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20892 |
| Brand Name: | Blood Donor Set/Quad/46347 |
| Model: | |
| Affected: | filter type LCRD2 |
| Software version: | |
| Recalling Organisation: | MacoPharma Australia Pty Ltd BDO Auckland L8 120 Albert St Auckland 1010 |
| Contact Information: | |
| Manufacturer: | Maco Pharma SA |
| Issue: | The manufacturer was informed of a number of fast filtrations on the reference NPT6280LA. This has caused an absence of effective filtration through the media of the LCRD2 filter. This defect is linked to a partial absence of the weld in the media caused by poor positioning (fold or gap) of the last polyester layer in the mattress. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 1/12/2016 |