Detail of Recall

Type of Product:Device
Medsafe Reference:20888
Brand Name:Laerdal Suction Unit
Model:
Affected:LCSU4, 300ml Complete Unit, LCSU 4, 800ml Complete Unit
Software version:
Recalling Organisation:Laerdal (New Zealand) Ltd
79 Apollo Drive
Rosedale
Auckland 0632
Contact Information:0800 251 400
Manufacturer:Laerdal Medical AS
Issue:The On/Off functionality in the LCSU 4 is controlled by electronic components. One of these components might be damaged. The damage occurs over time when the unit is connected to a power source (battery or external power source), but is not related to frequency of use.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/12/2016