Detail of Recall

Type of Product:Device
Medsafe Reference:20858
Brand Name:Siemens Artis Q, Q.zen, zeego, zee systems
Model:
Affected:Artis zee with SW versions VC21C, VD11 and Megalix Cat+ X-ray tubes - Lot numbers 10280959, 10094139, 10094141, 10094135, 10094137
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue: For Artis systems with software version VC21C, VD11, and Megalix Cat+ X-ray tubes, the semi-automatic focus switch may in cases of a special type of defect in foci small and micro (if present), not work as intended when pressing the footswitch multiple times. This special type of focus defect, that causes the failure of the semi-automated focus switch, becomes more likely as the service life of the X-ray tube increases.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/11/2016