Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20850
Brand Name: Roche Cobas - Homocysteine Enzymatic Assay
Model:
Affected:Cat No 05385415190 Lot 173114
Software version:
Recalling Organisation:Roche Consumer Health, Division of Roche Products (NZ) Limited
8 Henderson Place
Te Papapa
AUCKLAND
Contact Information:
Manufacturer:Roche Diagnostics GmbH
Issue:Following customer complaints, internal investigations have confirmed under recovery of QC and patient samples with Homocysteine reagent Cat No05385415190 Lot No 173114 by approximately 10-15% after 7 days exposure to 25oC. Correctly stored reagent may show low recovery after 3-4 months.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/11/2016