Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20841 |
| Brand Name: | Monaco RTP |
| Model: | |
| Affected: | V 5.00 and higher |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | Contours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm. If the CT data contain slices that occur at positions of xx.x50mm, there can be portions of the volume where there is a mismatch between the contour graphics shown on the screen and the contour data used during the optimization and calculation of dose. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/12/2016 |