Detail of Recall

Type of Product:Device
Medsafe Reference:20820
Brand Name:Brilliance iCT Ingenuity CT
Model:
Affected:Brilliance iCT / Brilliance iCT SP, Brilliance Big Bore Oncology / Brilliance Big Bore Radiology, Brilliance 6 / Brilliance 10 / Brilliance16/ Brilliance 16P / Brilliance 40 / Brilliance 64,Ingenuity CT / Ingenuity Core / Ingenuity Core128
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The patient support table top may fail to move in the horizontal direction when it is commanded to do so during CT scans, due to a potential mechanical linkage failure. The system cannot automatically detect this failure to abort the scan or to alert the user.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/11/2016