Detail of Recall

Type of Product:Device
Medsafe Reference:20813
Brand Name:VITROS Chemistry Products CRP Slides
Model:
Affected:Product Codes: 1926740 - 10758750000296, 8097990 - 10758750004355, GENs: 3, 11, 12, 30 through 36
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics
c/o Quigg Partners
Level 7, 36 Brandon Street
Wellington 6011
Contact Information:0800 402 899
Manufacturer:Ortho-Clinical Diagnostics Inc
Issue:Investigation confirmed that samples diluted (3X recommended dilution or higher) using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3 or a low CRP concentration sample had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The magnitude of bias increases as the dilution factor increases.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/11/2016