Detail of Recall

Type of Product:Device
Medsafe Reference:20798
Brand Name:Atrium Medical Thoracic Catheters
Model:
Affected:Trocar catheter code numbers 8408, 8410, 8412, 8416, 8420, 8424, 8428, 8432
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Atrium Medical Corporation
Issue:Atrium Trocar Catheters are packaged with a label containing an icon that shows more holes (eyelets) than the product actually contains. This graphical icon shows there are 6 side holes on the trocar catheter; however, the correct number of eyelets on the trocar catheter is 2 side holes.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/11/2016