Detail of Recall

Type of Product:Device
Medsafe Reference:20795
Brand Name:Elekta Monaco RTP
Model:
Affected:5.11 or 5.20
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:Dose can be incorrectly calculated when using the Reset function in Monaco®.
When creating 3D plans using Dose weighting mode, if the user rescales the plan, selects Reset, and then changes the wedge angle, the plan dose and monitor units (MU) should be returned to the original values before the rescale.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/11/2016