Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20795 |
| Brand Name: | Elekta Monaco RTP |
| Model: | |
| Affected: | 5.11 or 5.20 |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | Dose can be incorrectly calculated when using the Reset function in Monaco®. When creating 3D plans using Dose weighting mode, if the user rescales the plan, selects Reset, and then changes the wedge angle, the plan dose and monitor units (MU) should be returned to the original values before the rescale. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/11/2016 |