Type of Product:Device
Medsafe Reference:20753
Brand Name:Philips Achieva 3.0T and Ingenia 3.0T MR systems
Model:
Affected:Achieva 3.0T and Ingenia 3.0T MR systems with rear passive shielding
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The thickness and location of the shielding material must be installed according to the shielding requirements provided by the manufacturer. In cases where the requirements are not followed, the distance between the passive shielding rear wall and the magnet could be insufficient. An insufficient distance leads to stronger attraction forces between the magnet and the passive shielding. These stronger forces may lead to shearing of the magnet vibration pads.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/10/2016