Detail of Recall

Type of Product:Device
Medsafe Reference:20749
Brand Name:Siemens ADVIA Chemistry Calibrator
Model:
Affected:Lot numbers 680726, 680726A, 680726B, 680726C and 680726D
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The DBIL_2 Assigned Calibrator Values (ACV) may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for Direct Bilirubin measurement. Due to the proportional bias, the differences become greater as the concentration of DBIL_2 increases. Quality control (QC) material is also affected by this issue. Depending on the quality control concentrations and ranges used by the laboratory, quality control may not detect this issue.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/10/2016