Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20746 |
| Brand Name: | RayStation Version 5 Treatment planning |
| Model: | |
| Affected: | Version 5 |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | The manufacturer has reported an issue found with the display of dose computed on images other than the planning CT (auxiliary CT) when using multiple patient cases in RayStation 5. If a CT image set with the same Frame of Reference as the displayed auxiliary CT set exists in another case, the dose display may be incorrect |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/10/2016 |