Type of Product:Device
Medsafe Reference:20745
Brand Name:Abbott Diagnostics Architect Lactic Acid
Model:
Affected:31497UN14, 09353UN15, 45216UN15, 14596UN15, 37055UN15, 14782UN16
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Ireland
Issue:Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21). Which may have falsely depressed Lactic Acid results.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/10/2016