Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20736
Brand Name:Siemens Advia Centaur
Model:
Affected:Advia Centaur, Advia Centaur XP, Advia Centaur XPT
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed a change in correlation between serum and plasma specimen tubes with the ADVIA Centaur Vitamin D Total assay as compared to Instructions for Use (IFU). Newly evaluated data demonstrate a Deming regression slope between 0.93 and 0.95 for plasma samples as compared to serum samples whereas, the specimen collection comparison data provided in the IFU indicates a linear regression slope of up to 1.09
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/10/2016