| Type of Product: | Device |
| Medsafe Reference: | 20719 |
| Brand Name: | Covidien Devon Light Glove |
| Model: | 471711 31140208 31140216 and 31140257 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Covidien NZ Ltd Level 3, Building 5, Central Park Corporate Centre 666 Great South Road Penrose AUCKLAND 1051 |
| Contact Information: | 0508 489 264 |
| Manufacturer: | Covidien LLC |
| Issue: | Customers have reported that on rare occasions, the Devon™ Light Glove may split upon application to the Devon™ Light Handle Adapter. A subset of these splits reportedly resulted from a difficult application of the glove to the handle adapter. A split in the glove can potentially lead to a breach in the sterile barrier between the glove and the handle adapter. "After application, inspect the Light Glove for barrier integrity." |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/10/2016 |