Detail of Recall

Type of Product:Device
Medsafe Reference:20703
Brand Name:Laerdal Suction Unit (78xxxx)
Model:
Affected:7800033 LSU WITH RESUABLE CANISTER WITH NIMH BATTERY, 78001033 WITH ABBOT RECEPTAL CANISTER WITH NIMH BATTERY, 78002033 LSU WITH BEMIS CANISTER WITH NIMH BATTERY, 78003033 WITH SERRES SUCTION BAG CANISTER SYSTEM WITH NIMH BATTERY
Software version:
Recalling Organisation:Laerdal (New Zealand) Ltd
79 Apollo Drive
Rosedale
Auckland 0632
Contact Information:0800 251 400
Manufacturer:Laerdal Medical AS
Issue:This is a follow-up to the initial communication, to provide information on a long-term solution, with regards to the Laerdal Suction Unit (LSU) when used with the LSU NiMH Battery that has been stored or installed at low temperatures for a prolonged period.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/10/2016